RAE Systems UltraRAE 3000+ Portable Handheld Compound-Specific VOC Monitor - BLE, CSA, PID 9.8 eV lamp 1/2" with 10 RAE-Sep Tubes for Benzene, Rechargeable Li-Ion Battery, Universal Adapter, Monitor only - 059-D31D-000
Item #:
059-D31D-000
Brand:
RAE Systems
Share this product:
059-D31D-000 UltraRAE 3000+ Portable Handheld Compound-Specific VOC Monitor - BLE, CSA, PID 9.8 eV lamp 1/2" with 10 RAE-Sep Tubes for Benzene, Rechargeable Li-Ion Battery, Universal Adapter, Monitor only by RAE Systems
The UltraRAE 3000+ is one of the most advanced Compound-Specific Monitors on the market. Its Photoionization Detector's (PID) extended range of 0.01 to 10,000 ppm in VOC mode and 10 ppb to 200 ppm in benzene-specific mode makes it an ideal instrument for applications, from entry pre-screening during refinery and plant maintenance to hazardous material response, marine spill response and refinery down-stream monitoring.
- Dual detection mode for total benzene exposure assessment: 60-second snapshot or 15-minute STEL measurement
- High sensitivity to benzene provides a lower detection range for future benzene exposure limits
- Lower risk of false alarms through advanced speciation method
- Total VOC measurement mode with extended range of 0.01 to 10,000 ppm
FEATURES & BENEFITS
- Versatile VOC or Benzene-Specific modes
- Cost effective solution and low-cost RAE-Sep™ tubes
- Reflex PID Technology™
- Highly specific readings, combining a 9.8eV UV lamp and RAE-Sep™ benzene tube
- Built-in humidity sensor with Automatic temperature-controlled sampling time calculation
- See-through Sampling probe design for instant tubebreakthrough visibility
- Highly connectivity capability through multiple wireless module options
- Multi-language support
- Easy access to lamp, sensor and battery in seconds without tools
APPLICATIONS
- Confined space entry pre-screening during refinery and plant maintenance
- Hazardous material response
- Marine spill response
- Refinery down-stream monitoring
- Plant overhaul
MONITOR ONLY INCLUDES:
- UltraRAE 3000+ Monitor, Model PGM-7360
- Wireless communication module built-in
- Travel Charger
- RAE UV lamp and RAE-Sep™ Tubes
- Flex-I-Probe™ and short probe
- External filter
- Red rubber boot
- Alkaline battery adapter
- Lamp-cleaning kit
- Tool kit
- Lithium-ion (Li-Ion) battery with universal AC/DCcharger and international plug kit
- Soft leather carrying case
MONITOR WITH ACCESSORIES KIT:
- Hard transport case with pre-cut foam padding
- Charging/download cradle
- 5 porous metal filters and O-rings
- Organic vapor zeroing kit
- Gas outlet port adapter and tubing
OPTIONAL CALIBRATION KIT ADDS:
- Calibration gas, 34L, as specified
- Calibration regulator and flow controller
OPTIONAL GUARANTEED COST-OF-OWNERSHIP PROGRAM:
- 4-year repair and replacement warranty
- Annual maintenance service
Size | 10" L x 3.0" W x 2.5" H (25.5 x 7.6 x 6.4 cm) |
Weight | 26 oz (738 g) |
Sensors | Photoionization sensor with standard 9.8 eV or optional 10.6 eV or 11.7 eV lamps |
Battery | - Rechargeable, external field-replaceable Lithium-Ion battery pack - Alkaline battery adapter |
Operating Hours | 16 hours of operation |
Display Graphic | 4 lines, 28 x 43 mm |
Keypad | 1 operation and 2 programming keys, 1 flashlight on/off button |
Direct Readout | - VOCs as ppm by volume - High and low values - STEL and TWA - Battery and shutdown voltage - Date, time, temperature |
Alarms | - 95 dB buzzer at 12" (30 cm) and flashing red LED to indicate exceeded preset limits - High: 3 beeps and flashes per second - Low: 2 beeps and flashes per second - STEL and TWA: 1 beep and flash per second - Alarms latching with manual override or automatic reset - Additional diagnostic alarm and display message for low battery and pump stall |
EMC/RFI | Compliant with EMC directive (2004/108/EC) EMI and ESD test: 100MHz to 1GHz 30V/m, no alarm Contact: ±4kV Air: ±8kV, no alarm |
IP Rating | IP-65 unit running |
Datalogging | Standard 6 months at one-minute intervals |
Calibration | Two-point or three-point calibration for zero and span Reflex PID Technology™ Calibration memory for 8 calibration gases, alarm limits, span values and calibration dates |
Sampling Pump | - Internal, integrated flow rate at 400 cc/min - Sample from 100' (30m) horizontally or vertically |
Low Flow Alarm | Auto pump shut-off at low-flow condition |
Communication & Data Download | - Download data and upload instrument set-up from PC through charging cradle or using BLE module and dedicated APP - Wireless data transmission through built-in RF modem |
Wireless Network | Mesh RAE Systems Dedicated Wireless Network |
Wireless Range (Typical) | Up to 15ft (5m) for BLE EchoView Host: LOS > 660 ft (200 m) ProRAE Guardian & RAEMesh Reader: LOS > 660 ft (200 m) ProRAE Guardian & RAELink3 Mesh: LOS > 330 ft (100 m) |
Safety Certifications | US and Canada: CSA, Classified as Intrinsically Safe for use in Class I, Division 1 Groups A, B, C, D Europe: ATEX II 2G EEx ia IIC T4 IECEx CSA 10.0005 Ex ia IIC/IIB GbT4 |
Temperature | -4° to 122° F (-20° to 50° C) |
Humidity | 0% to 95% relative humidity (non-condensing) |
Attachments | Durable bright red rubber boot |
Warranty | 3 years for 10.6 eV lamp, 6 months for 9.8 eV lamp and 1 year for pump, battery, sensor and instrument |
Wireless Frequency | ISM license-free band. IEEE 802.15.4 Sub 1GHz |
Wireless Approvals | FCC Part 15, CE R&TTE, Others |
Radio Module | Supports BLE, RM900 or Bluetooth |
Gas Monitor | Range | Resolution | Response Time (T90) |
VOCs | 0 to 99.99 ppm 100 to 999.9 ppm 1000 to 9999 ppm |
0.010 ppm 0.1 ppm 1 ppm |
< 3 sec < 3 sec < 3 sec |
Benzene | 0 to 200 ppm | 0.010 ppm | < 60 sec |
Butadiene | 0 to 200 ppm | 0.010 ppm | < 60 sec |